Przeglądanie stron 40
39
INSERT/DELETE A STAGE
A
stage
is a section of a program that consists of a number of individual steps that can be cycled
(repeated) a defined number of times.
Insert a new stage
To insert a stage in a program, touch
Insert a
new stage
. A stage with one step will be
inserted.
Stage 1
Touch this to give the stage a name (optional).
Enter the name using the text entry screen.
1 cycle
Touch this to edit the number of cycles (repeats)
for the stage. Enter the required number of
cycles then touch
OK
.
Temperature
Touch the
temperature
button to edit the
temperature of the step and add any
increment/decrement or gradient features.
Time
Touch the
time
button to edit the hold time of
the step and to add any increment/decrement
time features.
Ramp rate
Touch the
ramp rate
setting to adjust the ramp
rate.
The range is 0.1ºC/s (minimum) to 5.0ºC/s
maximum for heating and 3.0ºC/s for cooling. The
value can be entered to 1 decimal place. Ramp rate
will affect the heating or cooling rate from the
previous step to the current step.
Once set, touch
OK
to return to the
Programming Screen.
Delete a stage
To
delete
a stage, touch the
stage area
to
highlight it and then touch
Delete
.