PRIME ELITE SATELLITE SPECIFICATION
TEMPERATURE
Block temperature range C or 20ºC below ambient to 100ºC
Block uniformity 0.3ºC
Temperature accuracy 0.2ºC
Gradient availability 10°C to 100°C
Maximum gradient 30ºC (16ºC with the 384-well block)
Minimum gradient 1ºC
Column uniformity with a 10ºC gradient 0.3ºC
Pre-run sample cooling Yes, 4ºC
Temperature set point adjustment 0.1ºC
HEATING/COOLING RATE
Maximum heating rate 5.0ºC/s
Minimum heating rate 0.1ºC/s
Maximum cooling rate 3.0ºC/s
Minimum cooling rate 0.1ºC/s
HEATED LID
Selectable heated lid temperature 35 to 115ºC or off
Pre-heat lid Yes
Warm up time from ambient <2 min
Over-temperature cut-out Yes
Regulated heated lid pressure Automatic
The heated lid is only operational if the block temperature is set above 35ºC
INTERCONNECTIVITY
Connection to Prime Elite Yes via USB port on rear of unit
Connection to PC control software Yes via USB port on rear of unit
DIMENSIONS
Height (mm) 312
Width (mm) 276
Depth (mm) 375
Weight (kg) 14
POWER
Voltage 100-230V
Power 1000W